Toric Lenses

Description:
Eyecryl TORIC are Aspheric Toric IOLs, which are designed to give excellent surgical outcome with predictability and visual performance on consistent basis.


Technical Specifications:

Material Hydrophobic Acrylic containing Natural Chromophore
Optic TypeSingle Piece, 360° Square Edge with Aspheric Optic
Optic Size6.00 mm
Overall Size 13.00 mm
Angulation
ACD5.28
Refractive Index1.48
Estimated A-Constants 118.5
Recommended Optical A-Constants SRK - T : 118.7
SRK - II : 118.9
Holl 1 Const SF : 1.66
HOFFER - Q ACD : 5.45
HAIGIS - a0:1.30, a1:0.40, a2:0.10
Barrett : 1.73
Hill-RBF : 118.70
Diopter Range +10.0 to +30.0 D
( with 0.5 D steps )
Implantation Site Capsular Bag
Sterilization Irradiation

Cylindrical Power Options

ModelCylindrical Power Recommended Correction Range of Corneal Astigmatism Correction
At IOL Plane At Corneal Plane*
HFY-05 1.00 D 0.68 D 0.25 - 0.86 D
HFY-10 1.50 D 1.03 D 0.87 - 1.25 D
HFY-202.25 D 1.54 D 1.26 - 1.75 D
HFY-303.00 D2.05 D1.76 - 2.25 D
HFY-35 3.75 D 2.57 D 2.26 - 2.75 D
HFY-404.50 D 3.08 D 2.76 - 3.25 D
HFY-50 5.25 D 3.60 D 3.26 - 3.75 D
HFY-60 6.00 D 4.11 D 3.76 D and above

Indications For Use

Eyecryl TORIC IOLs are intended to be implanted in the Capsular Bag of eye for visual correction of Aphakia& pre-existing Corneal Astigmatism in adult patients with or without Presbyopia, who desire improved uncorrected distance vision, reduction of residual refractive cylindrical error & increased spectacle independency for distance vision.


Features:


  • Predictable Astigmatic Correction
  • Excellent Rotational Stability
  • Reduced PCO with 360° Advanced Square Edge & tackiness


User friendly & accurate Eyecryl TORIC Calculator provides Eyecryl TORIC IOL model recommendation, Cylindrical Power & intended axis of IOL placement in the Capsular Bag.

Please Visit: www.biotechcalculators.com


 

Surgery Video - Toric Lenses

 

DIHFY600 IOLs can be manufactured in 5° angulation under special order requirement